Calibration and setup (SETUP) function:
After entering into calibration and setup (SETUP) function, the tester shall
display as follows:
----s e tu p ----
U n i t : m e t er
C a l i b r a t io n
Q u i t
UNIT: It is used to set up length unit and shifts between meter (Meter)
and feet (FT).
CALIBRATION is calibration function.(Detailed as related chapters hereinafter)
QUIT is used to return the main menu.
Dynamic calibration (CALIBRATION) function:
For an accurate measurement of cable length,the calibration operation should
be done as follows.
After entering into dynamic calibration function, the tester shall display as
follows:
C A L I B R A T IO N ?
N O Y E S
Insert same type cable of given length into M port, do not need insert
far-end recognizer, push key (Yes) to undertake measurement and
display the measured length (as shown below):
P i e a s e a d j us t ?
2 0 . 0 m
- o k +
At the moment, hold and key (-/+) to adjust the length to actual
given length and then push key to reserve calibration factor
and exit calibration function.It will display as follows if the cable
length being measured is too short(<10M) to remind the user to
change a longer cable for calibration:
C a b l e t o s h o rt !
c o h t i n n t . c ai
N o y e s
PAIR&L
10
Your excellent helper in cable test!